Appeal No. 1998-1789 Application No. 08/473,129 selected location along the length of the vessel, while at the same time not substantially occluding the vessel such that stenosis or thrombosis can occur. As manifested in claim 1, the invention comprises a catheter having tip immobilizing means on its distal end to “prevent the tip of the catheter from contacting a wall of the blood vessel without substantially obstructing fluid flow of blood through the blood vessel, such that catheter failure due to stenosis or thrombosis at the catheter distal end is reduced.” Similar limitations appear in all of the other claims. The examiner’s position is that the subject matter of all of the others except claim 26 is anticipated by the catheters shown in Figures 8-11 of Johnson. Johnson discloses a catheter within which are positioned a plurality of light-conducting fibers. The stated objective of the Johnson invention is to prevent the tips of the light- conducting fibers from contacting the wall of the vessel in which they are installed in order to avoid the creation of artifact by blood flow fluctuations and blood vessel wall reflectance (column 1, lines 32-59). In the embodiments shown in Figures 2-7, 10 and 11, Johnson accomplishes this objective 5Page: Previous 1 2 3 4 5 6 7 8 9 10 11 12 13 14 NextLast modified: November 3, 2007