Appeal No. 1998-2381 Application No. 08/304,725 flanking units are a pump assembly and at least one of a pump assembly, a parameter monitoring assembly, a display monitor or a controller. Like appellants, we fail to perceive any teaching or suggestion in the disclosures of Ford and Rota which would have motivated one of ordinary skill in the art to rearrange the discrete units of the APA in the claimed configuration. In fact, both Ford and Rota are lacking in any disclosure of separating discrete units into at least two flanking units with a central cavity therebetween adapted to receive or which is capable of receiving an oxygenator. The examiner's other assertion (Answer, page 4) that the claimed arrangement of known perfusion system components is a matter of design choice is not persuasive. In appellants' specification, at page 12, it is indicated that the centralized location of the oxygenator can reduce extracorporeal volume, improve the perfusionist’s reach of the system components and the perfusionist’s view of the operating 8Page: Previous 1 2 3 4 5 6 7 8 9 10 11 12 NextLast modified: November 3, 2007