Ex parte HOOD - Page 3




          Appeal No. 1999-2788                                                        
          Application No. 08/910,469                                                  


          tip.  In the examiner's view, this pointed tip in Davison is                
          at an outermost distal end of the blade body because,                       
               "[t]he point (at the end of the edge of the hook                       
               portion 62 shown in figure 8 of Davison et al.) is                     
               at the radially outermost point of the device.  In                     
               other words, no other portion of the device extends                    
               radially outward beyond this point.  This point is                     
               therefore at an outermost point of the body.  The                      
               point is also at a distal end of the body since it                     
               is located at the distal end portion of the body.                      
               The distal end portion of the Davison et al. body is                   
               considered to be the portion of the body which                         
               includes portions 60 and 64.  Although the extreme                     
               end of the sharped point is spaced proximally from                     
               the distalmost end point of the device, this is true                   
               for the blade of the invention as well.  The extreme                   
               end of the pointed tip 24 of the blade of the                          
               invention is spaced proximally from the distalmost                     
               end point of the device as seen in figure 1 of the                     
               application.  Even if the term "pointed tip" in the                    
               claims is considered to include the portion of the                     
               blade that has tapered sides that lead to the                          
               extreme end point (that is, the portion denoted by                     
               reference numeral 24 in figure 1), then even this                      
               pointed tip, including its base, is spaced                             
               proximally a small distance form the distalmost end                    
               point of the device as seen in figure 1 of the                         
               application."                                                          


          In the alternative, the examiner urges that "it would have                  
          been obvious that the tip described in col. 7, lines 34-36 of               
          Davison et al. is pointed for the reasons set forth above"                  
          (answer,                                                                    

                                          3                                           





Page:  Previous  1  2  3  4  5  6  7  8  9  10  11  Next 

Last modified: November 3, 2007