Ex Parte Sundar - Page 6




              Appeal No. 2004-0278                                                                Page 6                
              Application No. 09/626,362                                                                                


                     As shown in Figure 12, the sensor 52 further comprises a relative position                         
              detecting circuit 58 for detecting the position and orientation of the sense coil 54, both                
              relative to the source coil 53.  Provided outside the sensor 52 is a manipulator                          
              controller 59.  The controller 59 is designed to remote-control the surgical manipulator                  
              51 in accordance with the relative 3D position detected by the coils 53 and 54 of the 3D                  
              position sensor 52.  The 3D position sensor 52 and the manipulator controller 59                          
              constitute a remote-control system for controlling the manipulator 51.  The 3D position                   
              sensor 52 and the manipulator controller 59 constitute a remote-control system 55.                        


                     Mizuno teaches (column 14, line 35, to column 15, line 18) that:                                   
                            The remote-control system 55 can be applied to the first embodiment, i.e.,                  
                     the surgical manipulator system shown in FIG. 1--as illustrated in FIG. 14. As                     
                     shown in FIG. 14, two slave manipulators 3 and 4 are secured at their bases 7                      
                     and 8 to the right and left sides of an operating table 1, respectively. Two source                
                     coils 53, each functioning as a transmitting section, are secured to the right side                
                     of the operating table 1. Two sense coils 54, each functioning as a receiving                      
                     section, are held in the hands of a surgeon or attached to the head band the                       
                     surgeon wears. The first source coil 53 and the first sense coil 54 constitute a                   
                     first 3D position sensor 52a. The second source coil 53 and the second sense                       
                     coil 54 constitute a second 3D position sensor 52b.                                                
                            The first slave manipulator 3 is operated in accordance with the relative                   
                     3D positions of the coils 53 and 54 of the first position detector 52a. A task                     
                     coordinate system 62 is set, in addition to the base coordinate system 61 of the                   
                     first 3D position sensor 52a. The slave manipulator 3 is operated such that the                    
                     vector P representing the position which the sensor coordinate system 63 set for                   
                     the sense coil 54 and assumes in the task coordinate system 62 coincides with                      
                     the vector p representing the position which the TCP of the first slave                            
                     manipulator 3 takes in the base coordinate system 13 set for the first slave                       
                     manipulator 3. The task coordinate system 62 can be freely altered, by using the                   







Page:  Previous  1  2  3  4  5  6  7  8  9  10  11  12  13  14  15  Next 

Last modified: November 3, 2007