Ex parte HOBBS et al. - Page 4




          Appeal No. 96-3203                                       Page 4             
          Application No. 08/081,984                                                  


                 Appellants do not appear to dispute the examiner’s                   
          position that Heinke’s device “is capable of performing in                  
          the same manner as Applicant’s [sic, applicants’] device” as                
          set forth on page 3 of the answer. In support of this                       
          position, the examiner relies on Heinke’s disclosure that                   
          the coiled wire will “inhibit the blood flow to a high                      
          degree” thus inferring that some blood will flow past the                   
          coiled wire to filter the blood. We nevertheless cannot                     
          sustain the § 102(b) rejection of claim 21.                                 


                 Contrary to the examiner’s position as set forth on                  
          page 3 of the answer, Heinke does not disclose that the                     
          “largest loop is cylindrical” in the embodiment shown in                    
          Figure 6 or, for that matter, the other conical embodiment                  
          shown in Figure 5. Instead, the coiled wires shown in both                  
          of these embodiments are merely described as being                          
          “conical,” thus inferring that the coiled wire has a spiral                 
          form in which the diameter of the coiled wire continuously                  
          and progressively decreases from the base end to the apex of                
          the coiled configuration. In contrast, a loop or segment                    
          thereof would be required to have a constant diameter in                    







Page:  Previous  1  2  3  4  5  6  7  8  9  Next 

Last modified: November 3, 2007