Appeal No. 2002-1916 Application 09/332,772 internal diameter of sufficient size to accommodate a dilator or adapter (93) which is optionally positionable within the lumen of the catheter member (20). The adapter has an external diameter substantially equal to the internal diameter of the lumen of the catheter member (col. 16, lines 35-37), and the adapter (93) further includes a lumen defining an internal diameter for directly fittingly accommodating a relatively small diameter guidewire (95). Like the examiner, we are of the view that the lumen of the catheter member (20) in Donadio is of sufficient internal diameter so as to accommodate a first guidewire of a size larger than the size of the guidewire (95) accommodated by the dilator/adapter (93), and that this is all that the language of appellants’ claim 10 on appeal requires, because the first and second guidewires are not recited as positive elements of the claimed system. With regard to the optional positioning of the adapter (93) in the lumen of catheter member (20) in Donadio, we note that the patent indicates (col. 3, lines 5-7) that after placement of the catheter member (20) in a blood vessel the 5Page: Previous 1 2 3 4 5 6 7 8 9 10 11 NextLast modified: November 3, 2007