Ex Parte Pinchuk et al - Page 3

                Appeal 2007-1585                                                                             
                Application 10/383,268                                                                       
                Ground of Rejection                                                                          
                      Claims 25-28, and 30-35 stand rejected under 35 U.S.C § 103(a) as                      
                obvious in view of Fogarty.                                                                  
                      We affirm.                                                                             

                                               DISCUSSION                                                    
                Claims 25-28 and 34                                                                          
                      The Examiner contends that Fogarty teaches each of the three stent                     
                grafts claimed, but “that the claim language is not fully met in that the                    
                sealing end of graft module (92) (left side in figure 4) would have to be                    
                coupled to the left side of the graft module (94) in order to read on the claim              
                language.” (Answer 3.)                                                                       
                      However, the Examiner also finds that                                                  
                            Fogarty teaches that it was contemplated to vary the                             
                      diameters at the end or use a constant diameter section in the                         
                      system generally by the teachings associated with Figure 4. For                        
                      example, Fogarty teaches that the “length, cross-section, axial                        
                      bend angle, axial flexibility, resilient spring force . . . and the                    
                      like” can be varied to offer a variety of prosthetic character-                        
                      istics; see column 10, lines 52-56. Furthermore, barbs or                              
                      windings can be obviously utilized to attach the sections                              
                      together as was clearly contemplated by Fogarty; see Figures                           
                      3B to 3D and column 10, lines 16-33.  The barbs or windings                            
                      would make the outer diameter of module (92) larger than the                           
                      rest of the constant diameter section because of the presence of                       
                      the barbs or windings.                                                                 

                            On page 18, lines 7-17 of the specification, Appellant                           
                      states that the order of the stent-grafts could be changed or the                      
                      ends of the grafts could be reversed depending upon the                                
                      situation. Since it is clear that there was no criticality to the                      

                                                     3                                                       

Page:  Previous  1  2  3  4  5  6  7  8  9  Next

Last modified: September 9, 2013