Ex parte MAGGARD et al. - Page 15




          Appeal No. 1996-1813                                                        
          Application No. 08/117,453                                                  
          The use of two reference standards for calibration of the                   
          spectrophotometer would meet the “calibrate over a range”                   
          limitation of claim 1.  As stated above, one of ordinary skill              
          in the art would have been motivated to add the automated                   
          injection device to the spectrophotometric analysis system of               
          Maggard ‘785 in order to obtain a more efficient system.                    
                    Rejections over the Maggard WO‘762 reference                      
               Claims 1-18 have been rejected as unpatentable under 35                
          U.S.C. § 103 over combination of Maggard WO‘762, Inman and                  
          Dawes.  Applicants argue that the examiner has not                          
          established a prima facie case of obviousness.  Applicants                  
          position may be understood from the following excerpt from the              
          Brief and Reply Brief:                                                      
               The final rejection does not cite any patent which                     
               shows any automatic calibration of any instrument,                     
               much less the inventors’ automatic calibration with                    
               several fuels to cover a range of values for the                       
               property, e.g. octane, being measured.  Automatic                      
               none of the cited references suggest the periodic                      
               and automatic sampling from one or more hydrocarbon                    
               blends of different known composition, and the                         
               reporting of these calibration readings interspersed                   
               among the actual measurements taken by the analysis.                   
               (Brief page 18, lines 7-10).                                           
                                                                                     
          (Brief, page 16, line 4-7).                                                 
               When addressing the rejection of the claims 1-8 and 16-18              
          over the Maggard WO‘762 reference, applicants have not                      
                                         15                                           





Page:  Previous  8  9  10  11  12  13  14  15  16  17  18  19  20  21  22  Next 

Last modified: November 3, 2007