Ex Parte IKEDA - Page 11



          Appeal No. 2001-1512                                            11           
          Application No. 09/273,541                                                   

          the chuck.  White, in fact, assesses the state of the placement              
          of the wafer or separation of the wafer from the chuck by                    
          measuring a reduction in the electric current flow not by the                
          difference between the flow rate of the cooling gas and a                    
          predetermined value.                                                         
               Assuming, arguendo, that it would have been obvious to                  
          combine White’s method of assessing the wafer-chuck contact by               
          measuring the current with the apparatus of Tezuka, as held by               
          the examiner, Tezuka would still not disclose the claimed steps              
          of closing the exhaust valve so that all of the cooling gas                  
          passing through valve 16 is delivered to the gap between the                 
          wafer and the chuck.  In that regard, Tezuka only supplies what              
          appears to be a constant flow of cooling gas controlled by                   
          manipulating two valves while White monitors changes in the gap              
          between the wafer and the chuck by measuring the current flow to             
          the electrostatic chuck.  Accordingly, as the Examiner has failed            
          to establish a prima facie case of obviousness, the rejection of             
          claim 4, as well as claim 5 which depends therefrom, under 35                
          U.S.C. § 103 over Tezuka and White is not sustained.                         









Page:  Previous  1  2  3  4  5  6  7  8  9  10  11  12  13  Next 

Last modified: November 3, 2007