Ex Parte Callol et al - Page 9

                  Appeal 2006-3287                                                                                             
                  Application 10/022,996                                                                                       

                  45, 58 USPQ2d at 1065-66.   The court, focusing on the description of the                                    
                  coaxial lumen structure as the “basic sleeve structure for all embodiments of                                
                  the present invention contemplated and disclosed herein” concluded that the                                  
                  specification left “no doubt that a person skilled in the art would conclude                                 
                  that the inventor envisioned only one design for the catheters taught in                                     
                  SciMed's patents-an intermediate sleeve section containing two ... lumens                                    
                  arranged coaxially.” 242 F.3d at 1339-40, 58 USPQ2d at 1061-62.                                              
                          SciMed supports Appellants’ contention that one of ordinary skill in                                 
                  the art would understand the term “side by side” as used in the appealed                                     
                  claims as describing the type of configuration shown in Figure 34 of the                                     
                  Specification (see Reply 3 (citing Specification, p. 48, ll. 9-14)).  Cf. In re                              
                  Morris, 127 F.3d 1048, 1056, 44 USPQ2d 1023, 1029 (Fed. Cir. 1997)                                           
                  (approving the Board's definition of claim terms consistent with their                                       
                  definitions in CCPA cases).  Moreover, Appellants’ Specification, read as a                                  
                  whole, suggests that to properly function in accordance with the claimed                                     
                  method, the long and short balloons must have a “side by side” configuration                                 
                  of the type illustrated in Figure 34, rather than a coaxial configuration as                                 
                  shown in Lam, during the step of “mounting the stent on a catheter.”  See                                    
                  Teleflex, Inc. v. Ficosa N. Am. Corp., 299 F.3d 1313, 1326-27, 63 USPQ2d                                     
                  1374, 1381 (Fed.Cir. 2002)(specification and drawings provide context for                                    
                  construction of the claims).4                                                                                

                                                                                                                              
                          4 Compare Specification, p. 48, ll. 9-14 with p. 16, ll. 9-12 and p. 25,                             
                  ll. 24-25 (Figure 3 illustrates two stents implanted “side by side”),  p. 40,                                
                  ll. 10-13 (Figure 31 shows inner tubular members 172 and 174 within an                                       
                  inflation lumen 160 in a “side-by-side, radially spaced apart relation”) and                                 
                  p. 61, ll. 21-22 (In Figure 51, “the long balloon 500 is adjacent the short                                  
                                                              9                                                                

Page:  Previous  1  2  3  4  5  6  7  8  9  10  11  12  Next

Last modified: September 9, 2013