Ex Parte Westlund et al - Page 9

                Appeal  2007-0811                                                                              
                Application 10/128,997                                                                         

                the narrow veins.  We regard this as an express teaching or suggestion in the                  
                prior art for the use of the tapered flexible tip on the lead in Warman. We                    
                also regard the tip 24 disclosed by Williams as teaching a tapered tip for the                 
                same purpose.                                                                                  
                      Appellants argue that there is no teaching in Warman for the lumens                      
                to be isodiametric as claimed in claim 3.  On the contrary, we believe that                    
                the diameter of the outside of the lead of Warman is shown as uniform and                      
                Figure 2 is a typical cross section of the lead.  We believe that the Examiner                 
                is correct in concluding that there is no evidence to assume that the lumens                   
                are not isodiametric. We find that the Examiner has shown Warman                               
                discloses isodiametric lumens by a preponderance of the evidence.                              
                      With respect to the argued 120 degrees limitation of claims 7 and 20                     
                the Examiner states that it would have been obvious to arrange the                             
                electrodes in various formations, including being 120 degrees apart, in order                  
                to insure proper and efficient contact. Answer, page 6. We agree. It is                        
                apparent to us that the exact angle between the contacts is based partly upon                  
                anatomy, which is well known, and partly upon the geometry of the biased                       
                portion of the lead. One of ordinary skill would have found the 120 degree                     
                separation to have been obvious, inasmuch as the prior art clearly recognizes                  
                the importance of contact, and not only contact, but contact at the correct                    
                location. The level of skill with respect to an implantation pacemaker is                      
                quite high, both with respect to training and education and consequences of                    
                failure of the device. Thus, the exact angle between the electrodes is well                    
                within the skill in this art.                                                                  



                                                      9                                                        

Page:  Previous  1  2  3  4  5  6  7  8  9  10  11  12  Next

Last modified: September 9, 2013