Ex Parte Boutilette et al - Page 9

                 Appeal 2007-0845                                                                                      
                 Application 10/268,135                                                                                

                        We note that Jang’s drawings depict a gap between the angled                                   
                 structure and an inner surface of the wall 30.  However, “it is well                                  
                 established that patent drawings do not define the precise proportions of the                         
                 elements and may not be relied on to show particular sizes if the                                     
                 specification is completely silent on the issue.”  Hockerson-Halberstadt, Inc.                        
                 v. Avia Group Int’l, Inc., 222 F.3d 951, 956, 55 USPQ2d 1487, 1491 (Fed.                              
                 Cir. 2000); In re Wright, 569 F.2d 1124, 1127, 193 USPQ 332, 335 (CCPA                                
                 1977) (“Absent any written description in the specification of quantitative                           
                 values, arguments based on measurement of a drawing are of little value.”).                           
                        Appellants do not point to, and we do not see, any portion of Jang                             
                 describing precise proportions of the aperture and angled section of the                              
                 catheter, such that one of ordinary skill would conclude that the gap depicted                        
                 in the drawings would permit passage of a guide wire in both directions                               
                 along the guide wire lumen.  Rather, as discussed supra, Jang explicitly                              
                 states that the side port is configured to permit passage of a guide wire in a                        
                 distal direction from the side port 234, but not a proximal direction.                                
                        We also note that several of Jang’s drawings, for example Figure 3,                            
                 depict a guide wire passing underneath the angled tip structure.  However, as                         
                 discussed supra, the instant Specification discloses a guide wire exchange                            
                 procedure that “deflects the ramp 62 radially outward to allow the guide                              
                 wire 36 to pass thereunder” (Specification 9) (emphasis added).                                       
                        Thus, the fact that a guide wire may pass underneath Jang’s ramp                               
                 when deployed in a distal direction does not necessarily mean that the ramp                           
                 will fail to force a guide wire out of the guide wire lumen when the wire is                          
                 passed in a proximal direction.  This conclusion is consistent with Jang’s                            


                                                          9                                                            

Page:  Previous  1  2  3  4  5  6  7  8  9  10  11  Next

Last modified: September 9, 2013