Appeal No. 1997-3328 Application No. 08/226,605 Thereby, in accordance with 37 CFR ' 1.192(c)(7), claims 2-11 fall with claim 19. Thus, it follows that the decision of the examiner to reject claims 2-11 under 35 U.S.C. ' 103 is also reversed. The 35 U.S.C. § 103 rejection of claim 20. Appellants' claim 20 recites, 20. An electrophoresis device which comprises a supply channel, which contains a sample having an original sample composition, a drain channel, and an electrolyte channel, which contains an electrolyte buffer, wherein said supply and drain channels are each inclined with respect to the electrolyte channel, and which supply and drain channels intersect said electrolyte channel at a supply port and a drain port, respectively, such that a geometrically defined sample volume is defined by a section of said electrolyte channel located between said supply port and said drain port, which electrophoresis device further comprises a means for electrokinetically injecting a sample which reflects the original sample composition into said sample volume characterized in that said supply channel and said drain channel each have a resistance to flow with respect to said electrolyte buffer which is about 5% lower than the respective resistance to flow of said electrolyte channel. Appellants' describe (with respect to Fig. 3) that "[t]he resistance to flow of the supply and drain channel can be deminished by either reducing the length of the respective channels or by increasing their respective widths w" 10Page: Previous 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 NextLast modified: November 3, 2007