Interference No. 104,290 was conducted. LX-1173, 919. To perform the test, Fox placed the liver on a stainless steel ground plate" and connected the ground plate to the "Patient Plate" terminal of the electro surgical unit. LX-1173, 110. The four wire modified Dormier basket electrodes were loaded into the cannula. The cannula was inserted into the pig's liver. 1d. While holding the proximal end of the cannula with one hand, Fox advanced the proximal end of the modified Dormier basket with the other hand. The wire electrodes were advanced through the cannula and out into the pig's liver where they distended and assumed their deployed configuration. Id. Fox then energized the electrosurgery unit to supply rf current to the electrodes to ablate the tissue between the electrodes. Id. LX-1129 is a copy of Fox's lab notebook for June 23, 1993. LX-1173, T11. Fox placed a temperature probe in the center of the four electrodes to measure temperature under initial conditions and at the conclusion of the tests. Td. Fox recorded several tests with the electrosurgical instru ment operated at ever-increasing power. In all tests, the '0 The plate is elsewhere described as aluminum.Page: Previous 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 NextLast modified: November 3, 2007